Introduction |
This application note gives all the recommended data for the described analytical task, including the list of measured values based on certified reference samples. |
|
Alloy system Cu-Zn-Sn-Pb |
|
The Analytical problem |
Copper alloys will be often used in a wide range of composition for technical purposes. The most important alloys are brass with a high Zn-content and bronze with a high Sn-content. Other alloying elements can be Pb and Ni, Fe, Sb, Si < 0.5 % too. |
|
From the spectroscopical point of view the overlapping of Cu peaks with those of Zn and Ni has to be solved. But with spectra deconvolution this presents no problem. |
|
The Analytical instrument |
Measurements are made with the benchtop spectrometers |
|
 |
X-Ray ComPact or |
|
 |
X-Ray maXXi |
|
|
A low power tube is used in either instrument in excitation mode. A special collimator reduces the exciting beam down to 0.5 mm diameter. In this way, analysis of irregular or filigree samples will be possible. This is very important for jewellery manufacturers. |
|
Energy dispersive measurement is made with a proportional counter. Despite the limited resolution of this type detector, highly precise analysis is possible if the qualitative composition of the sample is known. Good statistical measurement guarantees high reproducibility. |
|
Max. Layer Thickness: approx. 50 µm |
|
Sample preparation |
The surface of the samples should be thoroughly cleaned free of contaminants. This will enable the irradiated area to be representative of the whole. |
|
Further preparation is not necessary as the very small excited area and the use of normalised intensities reduce the geometric influence on the signal. |
|
The irradiated area must be homogeneous. Non-homogenities can produce insufficient reproducibility of measurement. |
|
Excitation conditions |
| Voltage |
40 kV |
| Power |
approx. 15 W |
|
|
Program Parameters |
| Program |
SRX Vers. 4.104 |
| Evaluation program |
ALLOY.EXE |
| Measuring time for references |
180 s |
| Measuring time for analysis |
120 s |
| Normalisation parameter |
1 |
|
|
Parameter |
Au |
Cu |
Au |
Cu |
| Deconvol |
ON |
ON |
ON |
ON |
| Cmin |
55 |
0 |
0 |
0 |
| Cmax |
95 |
45 |
15 |
10 |
| ROI Int min |
130 |
140 |
420 |
210 |
| ROI Int max |
149 |
160 |
450 |
230 |
| ROI Decon min |
100 |
100 |
100 |
100 |
| ROI Decon max |
470 |
470 |
470 |
470 |
|
|
Reference samples |
The calibration was divided into two ranges - one for brass with relatively high Zn-content and small Sn-concentration and the other for bronze with relatively high Sn-contents and small Zn-concentrations. |
|
For approximately every 3 - 5 % one reference sample is necessary. For every calibration range analysed here, about 7 reference samples were used.
|
|
Analytical results |
Limits of detection: |
The limits of detection for elements that overlap with Cu (Ni, Zn) are approximately 0.5 %, and for the other elements (Pb, Fe, Sb) 0.2 - 0.3 %. |
|
Brass |
|
Accuracy |
Calibration table |
|
Cu |
Zn |
Sn |
Given |
Measured |
Given |
Measured |
Given |
Measured |
89.80 |
90.01 |
8.58 |
8.37 |
0.00 |
0.00 |
89.50 |
89.27 |
10.4 |
10.26 |
0.00 |
0.04 |
82.05 |
81.46 |
17.92 |
18.18 |
0.00 |
0.02 |
69.50 |
69.56 |
30.50 |
30.25 |
0.00 |
0.00 |
64.44 |
64.07 |
33.22 |
32.30 |
1.29 |
1.06 |
63.47 |
63.38 |
29.53 |
29.54 |
0.90 |
0.91 |
61.42 |
61.48 |
34.60 |
35.46 |
0.78 |
0.89 |
60.60 |
61.09 |
39.30 |
38.83 |
0.00 |
0.00 |
|
|
Error of fit |
For this calibration the following error of the fit was calculated [in Wt %]: |
|
|
|
Reproducibility |
Repeated measurements (120 s Measuring time) on the same sample gave the following results: |
|
Cu |
Zn |
Sn |
| SDConc [2 σ in Wt%] |
0.22 |
0.22 |
0.45 |
| RSDStatistic [2 σ in %] |
0.6 |
1.0 |
6.0 |
|
|
Bronze |
|
Accuracy |
Calibration table |
|
Cu |
Sn |
Pb |
Given |
Measured |
Given |
Measured |
Given |
Measured |
94.47 |
94.48 |
5.52 |
5.51 |
0.01 |
0.00 |
90.41 |
90.37 |
9.58 |
9.62 |
0.01 |
0.13 |
89.80 |
89.76 |
0.00 |
0.03 |
1.68 |
2.06 |
83.30 |
83.29 |
7.08 |
7.10 |
6.12 |
5.57 |
78.90 |
78.81 |
10.50 |
10.49 |
8.66 |
8.99 |
|
|
Error of fit |
For this calibration the following error of the fit was calculated [in Wt %]: |
|
|
|
Reproducibility |
Repeated measurements (120 s Measuring time) on the same sample gave the following results: |
|
Cu |
Sn |
Pb |
| SDConc [2 σ in Wt%] |
0.45 |
0.22 |
0.95 |
| RSDStatistic [2 σ in %] |
0.5 |
3.5 |
3.5 |
|
|