Introduction

This application note gives all recommended data for the described analytical task, including the list of measured values based on certified reference samples.

Alloy system       Ag-Cu

The Analytical problem

Analysis of Silver alloys is an important task in the jewellery industry for the determination of alloy composition and monetary value of alloys. There are no problems from the analytical point of view, because the X-Ray lines used have no overlapping.

The Analytical instrument

Measurements are made with the benchtop spectrometers

X-Ray ComPact or

X-Ray maXXi

A low power tube is used in either instrument in excitation mode. A special collimator reduces the exciting beam down to 0.5 mm diameter. In this way, analysis of irregular or filigree samples will be possible. This is very important for jewellery manufacturers.

Energy dispersive measurement is made with a proportional counter. Despite the limited resolution of this type detector, highly precise analysis is possible if the qualitative composition of the sample is known. Good statistical measurement guarantees high reproducibility.

Max. Layer Thickness: approx. 100 µm

Sample preparation

The surface of the samples should be thoroughly cleaned free of contaminants. This will enable the irradiated area to be representative of the whole.

Further preparation are not necessary as the very small excited area and the use of normalised intensities reduce the geometric influence on the signal.

The irradiated area must be homogeneous. Non-homogenities can produce insufficient reproducibility of measurement.

Excitation conditions

Voltage 40 kV
Power approx. 15 W

Program Parameters

Program SRX Vers. 4.104
Evaluation program ALLOY.EXE
Measuring time for references 120 s
Measuring time for analysis 120 s
Normalisation parameter 3

Parameter

Au

Cu

  Deconvolution

On

On

  Cmin

60

0

  Cmax

100

40

  ROI Int min

364

130

  ROI Int max

400

150

  ROI Decon min

300

90

  ROI Decon max

470

180

Reference samples

For approximately every 5 - 10 % one reference sample is necessary. For the concentration range analysed here, five reference samples were used.

Attention:

Because of the lower cost value of silver, alloys often contain a greater than normal percentage of silver. Hallmarks should not therefore be used as analyticalal references!

Analytical results

Accuracy

Calibration table

Ag

Cu

Given

Measured

Given

Measured

100.00

100.00

0.00

0.00

97.00

97.19

3.00

2.81

93.50

93.37

6.50

6.63

87.00

86.62

13.00

13.38

83.50

83.71

16.50

16.39

62.80

62.95

37.20

37.05

Error of fit

For this calibration the following error of the fit was calculated 0.05 Wt %

Reproducibility

Repeated measurements on the same sample (93 Wt % Ag) gave the following results:

SD (Ag-Concentration): 0.05 Wt %
RSD (Ag-Intensity [1*s]): 0.33 %
RSD (Statistic Ag-Intensity): 0.27 %

«« Back to Applications