|
Introduction |
The traditional methods determining Gold in precious metal alloys, preferably jewellery alloys, are carried out by the Touchstone test or by the Cuppellation. But these methods are destructive and time consuming. Therefore since many years the X-ray Fluorescence Spectrometry has been introduced to solve this problem. It is the power means for precise determination of alloy composition and concentrations of Au or Ag. |
|
Alloy system Au-Ag-Cu |
|
The analytical instrument |
The instrument Xray ComPact 5/PIN offered by Roentgenanalytik Messtechnik GmbH has been especially developed for the analysis of jewellery alloys made of gold, silver or platinum. |
|
A low power tube is used in either instrument in excitation mode. A special collimator reduces the exciting beam down to 0.6 mm in diameter. In this way a local analysis of several parts of the jewellery object and its filigree zones with irregular geometry, as usual for jewellery manufacturing, can be achieved. |
|
Energy dispersive measurement is made with a PIN-diode. The high resolution of this type of detector allows to analyse trace elements in samples with complex composition. Good statistical measurement guarantees high reproducibility. |
|
Sample Preparation |
|
The surface of the samples should be thoroughly cleaned free of contaminants. This will enable the irradiated area to be representative of the whole. |
|
Further preparation is not necessary as the very small excited area and the use of normalized intensities reduce the geometric influence of the signal. |
|
The following figure 1 shows a typical spectrum of AuAgCu alloy. |
|

|
Figure 1: spectrum of AuAgCu alloy |
|
Excitation conditions |
| Tube |
W-Microfocus-glass-window |
| Voltage |
40 kV |
| Power |
approx. 30 W |
| Collimator |
0.4 mm in diameter |
| Measuring time for analysis |
60-120 s |
|
|
Program Parameters |
| Program |
XMaster Vers. 2.0.2.3 |
| Application |
AuAgCu |
|
|
Analytical Features |
|
Limits of detection |
Traces in alloys of about 0.1% can be analysed.
In most cases this sensitivity is more than sufficient for jewellery analysis.
|
|
Calibration ranges |
The spectrometer is calibrated for gold, silver and platinum alloys. For calibration certified reference samples are used. Calibration is valid for the concentration ranges shown in table 1. |
|
Alloy System |
Au |
Ag |
Cu |
Zn |
Pd |
Ni |
Gold |
30 - 100 |
0 - 70 |
0 - 40 |
0 - 10 |
0 - 30 |
0.20 |
|
Table 1: Calibrated concentration ranges |
|
Accuracy of analysis |
An accuracy of ± 0.1% can be achieved within an analysis time of 2 min; this corresponds to 0.05 Carat. For all other admixtures the accuracy is in the same range. The accuracy of the X-ray spectrometric determination of Gold shall be validated by certified standards of jewellery alloy samples. The following table 2 enables to compare X-ray measured values with these of Gold standards. |
|
|
Alloy System |
Ag |
Ag |
Cu |
given |
is |
given |
is |
given |
is |
Au1 |
33,32 |
32,40 |
66,59 |
67,59 |
|
|
Au2 |
33,35 |
33,38 |
44,65 |
45,19 |
21,98 |
21,42 |
Au13 |
91,67 |
91,67 |
2,76 |
2,98 |
5,28 |
5,35 |
Au15 |
98,60 |
98,78 |
|
|
1,40 |
1,21 |
Au16 |
99,99 |
100,0 |
|
|
|
|
|
Table 2: Analysis results. |
|
Reproducibility |
For determination of reproducibility repeated measurements with a measuring time of 60s ahs been realised with 10 min on a sample with the following composition [%o]- Au-Ag-Cu: 916.7 – 27.6 52.8 |
|
No of meas. |
Au |
Ag |
Cu |
1 |
916,6 |
29,8 |
53,6 |
2 |
917,4 |
29,1 |
53,4 |
3 |
917,0 |
30,0 |
53,0 |
4 |
916,8 |
30,0 |
53,1 |
5 |
917,1 |
29,3 |
53,6 |
6 |
917,0 |
29,9 |
53,0 |
7 |
915,7 |
30,2 |
54,1 |
8 |
916,3 |
29,5 |
54,1 |
9 |
916,0 |
30,3 |
53,6 |
10 |
917,1 |
29,7 |
53,1 |
Mean |
916,7 |
29,8 |
53,5 |
Error rel [%] |
0,05 |
1.34 |
0,75 |
SD [%] |
0.5 |
0.4 |
0,4 |
|
|
Conclusion |
The analytical results demonstrate that the standardless X-ray fluorescence analysis is a most important alternative methods for the analysis of jewellery alloys. The X-ray spectrometer ComPact 5/PIN allows the determination of Gold in /jewellery alloys with high accuracy and high precision. In comparison with the conventional touch method, the analytical accuracy is improved by a factor of 10. Not only is the X-Ray fluorescence analysis non-destructive, it does not need high operator skill and highly accurate results are quickly obtained. Extremely narrow collimated X-Ray beams enable even the fines filigree jewellery to be analysed. |
|
|
|
|